C-Arm Refurbishing Process
- Run serial number with OEM for necessary upgrades, recalls, and revisions (software and firmware included)
-
Verify presence, function, and performance of internal components including, but not limited to:
- High voltage x-ray generator
- High voltage cables
- X-ray tube
- Image intensifier
- Vertical lift assembly
- Steering mechanism
- Power supplies
- Monitor displays
- Repair, rebuild, or replace any internal components listed above that do not meet OEM specifications
- Perform complete system disassembly
- Disinfect all components
- Sand and repaint all previously painted surfaces using a 4-coat process as performed by OEM
- Reassemble refurbished components
- Inspect and replace wiring harnesses as necessary
- Lubricate all bearings and replace as necessary
- Clean and calibrate all locking and braking systems
- Inspect, clean, and lubricate high voltage cables
- Install new batteries
- Calibrate image quality to ensure optimum performance
- Calibrate all other systems according to OEM specifications
- Perform comprehensive manufacturer PM service prior to staging
- Configure system to customer's specified parameters
- Stage system for optimum performance and reliability
- Label machine and components to specified requirement